LifeUltra Autoclave
Powerful. Unique. Revolutionary.
Powerful. Unique. Revolutionary.
The LifeUltra tabletop steam sterilizer stands out as the first system with a built-in water distiller, eliminating the need for distilled water while supporting a direct facility water supply. It delivers high efficiency with a true 25L chamber, up to 92% volumetric capacity, and optimized sterilization cycles that reduce water and power consumption.
Designed for reliability and safety, it features robust chamber construction, multiple locking mechanisms, and compliance with major industry standards like ANSI, IEC, FDA guidelines. With eco-friendly performance, integrated descaling, and a strong warranty backed by U.S.-based manufacturing and support, its offers a sustainable and user-friendly sterilization solution.
Functions
Sterilization programs – B275 Universal, B250 Universal, S275 Universal, S250 Universal
Test programs – Bowie & Dick, Vacuum Function
Other programs – Descaling Water Distiller
Electric
Voltage input – 115 V ~ 125 V AC – 60 Hz
Current drawn – 18 A – 1 Ø
Nominal power rating – 1950 VA
Thermal circuit breakers – 392°F Distiller and Chamber, 482°F Steam Generator
Insulation class – Class A
Air, Water and Steam
Bacteriological air filter – 0.027 micron Porosity
Municipal feed water – ≤500 ppm TDS at 68°F
Manual feed water – ≤7 ppm TDS at 68°F
Feeding water tanks – 3L
Recirculating water tanks – 2L
Water distilling rate – 500
Dimensions and Weights
Device dimensions – 24″ width x 20″ height x 25″ depthDevice weight – 198 lb (unladen) | 230 lb (with full tanks and trayed chamber)
Chamber dimensions – 10.8″ dia x 15.8″ depth
Chamber volume – about 25 L
Chamber loading – 15 lb unwrapped loads, 10-12 lb wrapped loads, 2.2 lb wrapped delicates, 7 wrapped handpieces
Features
Display interface – 4.8” graphic monochrome 240 x 128 pixel backlit LCD
Control interface – 3 navigation buttons and 1 selection button
Hard copy output – Thermal printer with 21⁄4” spill-proof paper roll holder and an integrated manual cutter
Environment
Installation category – Cat. II (EN 61010-1 : 2010 + A1 : 2019)
Operating conditions – 60°F ~ 95°F ambient temperature, 20% ~ 80% non-condensing RH, 10 air changes per hour (ACPH)